Page 62 - Catalog Instruments and Supplies
P. 62

62 SCP SCIENCE
     EasyPREP - Laboratory Automation

Automated BOD, CBOD and DO Analysis
BOD-200

The BOD-200 provides users with the capability to conduct BOD, CBOD         Features
and DO analysis. Analyze CBOD in conjunction with BOD to verify if non-
compliant BOD results are due to nitrification and to obtain results more   BOD, CBOD and DO unattended automated analysis
reflective of real effluent quality when under nitrifying conditions.       Optimize laboratory management by opting for
                                                                            a 5 or 7-day incubation periods
Biochemical oxygen demand, BOD, measures oxygen depletion which             Accommodates 300 ml reusable glass and/or disposable bottles
represents the breakdown of organic material by microorganisms.             and 60 ml glass re-usable bottles
                                                                            Barcoding option provides sample, bottle and rack tracking
Carbonaceous BOD, CBOD, measures oxygen depletion from the                  throughout the process
breakdown of organic material by microorganisms in which the                Auto wash station prevents cross contamination
contribution from nitrogenous bacteria has been suppressed.                 User-friendly software, LIMS compatible
                                                                            Compliant with laboratory standards:
The BOD-200 operates with an Optical Dissolved Oxygen Probe,
providing accurate and reproducible results. The Optical Probe requires         ISO 5815-1 | EN-1899-1 | EN-1899-2 | EN-5814 (2012) | EPA 405.1
no membrane maintenance or electrolyte re-filling and is characterized
by its stable response, minimal drift and fewer single point-calibrations.                                           designed for
                                                                                                                   State, Municipal,
Full Process Automation provides unattended:                                                                       and commercial
                                                                                                                    environmental
•	 Automated Seed Addition
                                                                                                                      testing labs
•	 Automated Inhibitor Addition

•	 Automated Sample Dilution

•	 Automated Sample Homogenization

•	 Automated Data Acquisition, Calculation and Storage

specifications

               Probe

Range          0-50 mg/L.

Accuracy       0 to 20: ±0.1mg/L or ±1% reading,
               whichever greater
Resolution
Response Time  20-50 mg/L, ±15% reading

               0.01 mg/L

               T95=22 seconds with stirring

www.scpscience.com | e-mail: [email protected] | phone: +1 (514) 457-0701