Page 64 - Catalog Instruments and Supplies
P. 64

64 SCP SCIENCE
     EasyPREP - Laboratory Automation

Automated Chemical Oxygen Demand Analyzer
cOD-200

This, all-in-one, unattended COD analysis system automates the         COD 200 offers the latest in solid state optics including LED
mixing, heating, cooling and analysis of COD samples.                  sources and Si detectors for all wavelengths stated in world wide
                                                                       government regulations.
COD tubes are placed in a digestion block for sample digestion,
followed by shaking and cooling. They are then transferred to the      In COD analysis, the presence of turbidity can cause deviations of up
spectrophotometer where sample are analyzed. The measurement           to 30% from the real sample value. This capability allows the end
activity and concentration calculations are displayed in real-time on  user to confidently analyze samples knowing that the turbidity level
the screen.                                                            is adequate for COD analysis.

                                                                                designed for
                                                                              State, Municipal,
                                                                              and commercial
                                                                               environmental

                                                                                 testing labs

Features                                                               Features

Sample capacity of 217 COD tubes                                       Complies with ISO 15705, ISO 8467 and EPA 410.4
                                                                       Digestion block with a heating temperature range from
Turbidity detection capability                                         ambient to 180° C
Measure samples 0.5 mg/l to 1500 mg/l                                  User friendly software with graphical display of the
Fully automated system from sample pick up to output of results        samples being processed
Includes: digestion, sample time tracking, shaking, analysis, data     LIMS compatible output file. Sample ID’s are saved for full
transposition and storage                                              traceability
                                                                       Computer system requirements can be found on page 63

www.scpscience.com | e-mail: [email protected] | phone: +1 (514) 457-0701