Page 19 - Catalog Instruments and Supplies
P. 19

SCP SCIENCE 19
                                                                                               Digestion & Evaporation Systems

                                                                                               Method Development Guide
                                                                                               HIGHLY REACTIVE samPLES

HIGHLY REACTIVE EXAMPLES                                   Highly
i.e. Animal Feeds, Waste Oils*,                           Reactive
                                                          Samples
     Polymers, Unknown Samples

     * Not to be used with HNO3/HClO4

                                                 Choose Acid Combination

PlasmaPURE H2SO4 Acid                                                           PlasmaPURE H2SO4/HNO3 Acids
- Will oxidize, hydrate or sulfonate organic compounds                          - First add H2SO4 to char sample:
- Works best above 230°C
- Glass or Quartz sample tubes                                                    Sample should be above 200°C
                                                                                - Then add HNO3 to digest the sample:
PlasmaPURE HCl Acid
- Strong non-oxidizing mineral acid                                               Sample should be between 80°C and 120°C
- Sample should be below 80°C                                                   - Glass or Quartz sample tubes
- DigiTUBE, Glass, or Quartz sample tubes, Watch Glasses
                                                                                PlasmaPURE HNO3/HClO4 (should always be used with
PlasmaPURE HNO3 Acid                                                            HNO3) Acids
- Strong oxidizing mineral acid                                                 - Very strong combination
- Sample digested between 80°C and 120°C
- DigiTUBE sample tubes, Watch Glasses                                          - Sample should digest between 145°C and 200°C

PlasmaPURE HF Acid                                                              -  DMiugsiTt UusBeEH, GCllaOs4s,fuomr eQhuoaordtz sample tubes, Watch Glasses
- Strong non-oxidizing acid                                                     -
- Ideal for samples such as SiO2
- Sample should digest below 95°C                                               PlasmaPURE HNO3/HCl Acids
- Teflon® sample tubes                                                          - Very strong oxidizing media
                                                                                - Reacts upon mixing - forms brown fumes upon mixing
                                                                                - Samples should be digested between 95°C and 110°C
                                                                                - DigiTUBE, Glass, or Quartz sample tubes, Watch Glasses

Below 180 °C                                              Choose System            Above 180 °C
DigiPREP MS, Jr. or LS, h, MINI &                         Choose Weight            DigiPREP HT & Touch Screen Controller
Touch Screen Controller                                                            Use: "Time to Temperature"

Use: - Multiple "Time to Temperature"                                                      "Time at temperature"
       - Multiple "Time at Temperature"
       - DigiPROBE to control internal                                                     Select 100 ml or 250 ml tubes
        sample temperature

Below 2 grams                                                                      Above 2 grams
 - Add 5 ml total acid at starting point                                            - Add 10 ml total acid at starting point

 - Always add acid in small aliquots                                                - Always add acid in small aliquots

                                                          Digest Sample

                                             NO           Is Digestate clear?*                 * Not sole criteria for a complete digestion.

                   Is there precipitate?                             YES          YES            Is Digestate clear?*
                        YES                                                                                  NO
                                                               Digestion        NO
NO Dilute the sample with water and                            Complete                   YES  Review the initial acid
            see if the precipitate remains.                                                           combination

            Is the digestate clear, but acid                                                    Reduce the initial acid
                   concentration high?                                                                concentration
   14   15   16   17   18   19   20   21   22   23   24