Page 21 - Catalog Instruments and Supplies
P. 21

SCP SCIENCE 21
                                                                        Digestion & Evaporation Systems

DigiPLATE is a 24 x 10 ml Automated Well Plate Digestion                                                    EasyPREP
System capable of multiple liquid handling operations, digestion,                                              DigiPLATE
and then filtering small volume samples in a compact, standard
well-plate configuration. Digestion with internal exhaust, dispensing                                                                     DigiPLATE
and filtration are built into a fully automated system controlled by a
stand-alone computer.                                                   DigiPLATE with filtration Module

Standard features include the capability to add acids prior to                       Ordering Information
digestion, sample digestion, sample cooling, adding additional
reagents and/or internal standard(s), normalizing samples to a fixed        Qty Catalog No.
volume and aliquoting samples into a second plate. If equipped                1 010-430-001
with an optional Filtration Module, the system will then take final          12 010-430-010
digested samples and filter them simultaneously through a 0.45 or            12 010-430-020
1.0 micron filtration membrane plate.                                         6 010-430-030
                                                                              1 010-430-003
With a small sample size, less acid is used, less waste is generated          1 010-430-002
and less corrosive gases are produced. Fumes are easily handled by            1 010-400-008
the internal exhaust system.
DigiPLATE heats samples from ambient to 180 °C in less than
10 minutes. Software includes unlimited time to temperature and
time at temperature digestion methods, programmed cooling times,
ability to store and recall unlimited sample methods, set reagent
addition volumes and normalization parameters, set aliquot volumes
and filtration times. The software manipulates 24 x 10 ml well-
plates through each station (Digestion, Reagent Addition/Cooling
and Filtration) allowing for completely unattended operation.

                       PC minimum requirements:

 Windows® XP, Vista, 7 or 8. Processor 1 GHz or superior, 1GB RAM,
 HD minimum 1 GB free space and USB port.

                                  Description

 DigiPLATE 110/230V with software
 Well plate 10 ml, 24 position
 Watch glass plate
 Filter plate 0.45 Micron Hydrophilic
 Filter Module option
 DigiPLATE with Filtration Module